What to do if...

83

6

What to do if...

6.1

pH and ORP measurement

Error message,

OFL or UFL

(measuring range

exceeded or undercut)

Error message, Error

(calibration error)

Cause

Remedy

Electrode:

– Air bubble in front of the

diaphragm

– Remove air bubble

– Air in the diaphragm

– Extract air or

moisten diaphragm

– Gel electrolyte dried out

– Replace electrode

Test sample

– The pH value lies outside the

measuring range

– not possible

Cause

Remedy

Electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

– Moisture in the plug

– Dry plug

– Not enough electrolyte

– Top up electrolyte

– Electrode obsolete

– Replace electrode

– Electrode broken

– Replace electrode

– Socket damp

– Dry socket

Calibration procedure:

– Incorrect solution temperature

(without temperature sensor)

– Set up correct temperature

– Incorrect buffer solutions

– Select buffer solutions

suitable for the calibration
procedure

– Buffer solutions too old

– Use only once.

Note the shelf life