Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 35

Multi 3500i
Operation
33
Calibration evaluation
After calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worse evaluation of both is taken into account. The evaluation appears
on the display and in the calibration record.
Preparatory activities
Perform the following preparatory activities when you want to calibrate:
Display
Calibration
record
Asymmetry
[mV]
Slope
[mV/pH]
+++
-15 ... +15
-60.5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the electrode according
to the electrode operating
manual
Error
Error
< -30 or
> 30
< -62 or
> -50
Perform error elimination ac-
cording to chapter 6 W
1
Connect the pH electrode to the measuring instrument.
The pH/ISE measuring window is displayed.
2
Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature, if you
measure without a temperature sensor.