2 ise measurement, Ise measurement – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 87

Multi 3500i
What to do if...
85
6.2
ISE measurement
Error message OFL
Obviously incorrect
measured values
Error message, Error
(invalid
calibration)
Warning [TempErr]
Warning [ISEErr]
Cause
Remedy
– Measuring range exceeded
– Dilute test sample
Cause
Remedy
– Electrode not connected
– Connect electrode
– Cable broken
– Replace electrode
Cause
Remedy
Ion-sensitive electrode:
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace electrode
– Electrode unsuitable for the
range to be measured
– Use a suitable electrode
– Socket damp
– Dry socket
Calibration procedure:
– Wrong sequence of standards
for three point calibration
– Select correct sequence
– Calibration standards do not
have the correct temperature
(max.
±
2 °C temperature
difference)
– Adjust the temperature of
the calibration standards
Cause
Remedy
– Temperature difference
between measurement and
calibration greater than 2 K.
– Adjust the temperature of
the test sample
Cause
Remedy
– Electrode voltage outside
calibrated range
– Recalibrate