10 reset, 1 resetting the sensor settings – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 80

Operation
Multi 3500i
78
4.10
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.10.1 Resetting the sensor settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH
The following settings for pH measurements are reset to the default
settings with the Reset function:
Oxi
The following settings for DO measurements are reset to the default
settings with the Reset function:
Setting
Default settings
Calibration type
TEC
Cal. interval
7 d
Unit for slope
mV/pH
Measured variable
pH
High resolution
On
Asymmetry
0 mV
Slope
-59.16 mV
Temperature, manual
25 °C
Setting
Default settings
Cal. interval
14 d
Comparison meas.
Off
Measured variable
DO concentration
Relative slope (S
Rel
)
1.00
Salinity (value)
0.0
Salinity (function)
Off