Technical data

Multi 3500i

92

7.2.3

Dissolved oxygen

Measuring ranges,

resolution

Note: The values quoted in brackets apply especially for the
DurOx 325 sensor.

Variable

Measuring range

Resolution

DO concentration
[mg/l]

0 ... 20.00 (0 ... 20.0)
0 ... 90.0 (0 ... 90)

0.01 (0.1)
0.1 (1)

Saturation [%]

0 ... 200.0 (0 ... 200)
0 ... 600

0.1 (1)
1

O

2

partial pressure

[mbar]

0 ... 200.0 (0 ... 200)
0 ... 1250

0.1 (1)
1

T [°C]

0 ... 50.0

0.1

Accuracy (± 1 digit)

Variable

Accuracy

DO concentration
[mg/l]

± 0.5 % of measured value
at ambient temperature + 5 °C ... + 30 °C

Saturation [%]

± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature

O

2

partial pressure

[mbar]

± 0.5 % of measured value
at ambient temperature + 5 °C ... + 30 °C

Correction functions

Temperature
compensation

Accuracy better than 2 % at 0 ... + 40 °C

Salinity correction

0 ... 70.0 SAL

Air pressure
correction

Automatic through integrated
pressure sensor in the range of
500 ... 1100 mbar