3 measuring the orp voltage, Measuring the orp voltage – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 30

Advertising
background image

Operation

Multi 3500i

28

4.4.3

Measuring the ORP voltage

Note
ORP electrodes are not calibrated. However, you can check ORP elec-
trodes using a test solution.

AutoRead

(drift control)

The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured vari-
able flashes until a stable measured value is available.

Criteria

With identical measurement conditions, the following applies:

1

Perform the preparatory activities according to section 4.6.1.

2

Submerse the ORP electrode in the sample.

3

Select the mV display with <M>.

U

157.0

mV

24.8

°C

Measured variable

Reproducibility

Response time

ORP voltage

better than 0.6 mV

> 30 seconds