2 automatic data storage at intervals, Automatic data storage at intervals – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 70

Operation
Multi 3500i
68
4.8.2
Automatic data storage at intervals
The storage interval (Interval) determines the chronological interval be-
tween automatic storage processes. Each storage process transmits
the current dataset to the interface at the same time.
Configuring the auto-
matic data storage func-
tion
Settings
You can configure the automatic data storage function with the follow-
ing settings:
1
Press the <STO> key for a long time.
The menu for automatic data storage appears.
Store with ID no.
1
Interval
30 s
Duration
180 min
Continue
00d03h00min
00d15h
34 min
<STO_>
Menu item
Possible
setting
Description
Store with ID no.
1 ... 999
ID number for the dataset
series.
Interval
5 s, 10 s, 30 s,
1 min, 5 min,
10 min,
15 min,
30 min, 60 min
Storage interval.
The lower limit of the stor-
age interval can be restrict-
ed by the number of free
storage locations.
The upper limit is restricted
by the storage duration.
Duration
1 min ... x min
Storage duration
Specifies after which time
the automatic data storage
should be terminated.
The lower limit of the stor-
age duration is restricted
by the storage interval.
The upper limit is restricted
by the number of free stor-
age locations.