2 automatic data storage at intervals, Automatic data storage at intervals – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 70

Advertising
background image

Operation

Multi 3500i

68

4.8.2

Automatic data storage at intervals

The storage interval (Interval) determines the chronological interval be-
tween automatic storage processes. Each storage process transmits
the current dataset to the interface at the same time.

Configuring the auto-

matic data storage func-

tion

Settings

You can configure the automatic data storage function with the follow-
ing settings:

1

Press the <STO> key for a long time.
The menu for automatic data storage appears.

Store with ID no.

1

Interval

30 s

Duration

180 min

Continue

00d03h00min

00d15h

34 min

<STO_>

Menu item

Possible
setting

Description

Store with ID no.

1 ... 999

ID number for the dataset
series.

Interval

5 s, 10 s, 30 s,
1 min, 5 min,
10 min,
15 min,
30 min, 60 min

Storage interval.

The lower limit of the stor-
age interval can be restrict-
ed by the number of free
storage locations.
The upper limit is restricted
by the storage duration.

Duration

1 min ... x min

Storage duration
Specifies after which time
the automatic data storage
should be terminated.

The lower limit of the stor-
age duration is restricted
by the storage interval.
The upper limit is restricted
by the number of free stor-
age locations.