Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 64

Operation
Multi 3500i
62
Temp. comp. (TC) /
Linear coeff.
0.000 ...
3.000 %/K
Coefficient of the linear tem-
perature compensation.
This setting is only available
when the linear temperature
compensation is set.
Measuring cell
Measuring cell used
Cal
Measuring cells the cell con-
stant of which is determined
by calibration in the KCL con-
trol standard solution.
Calibration ranges:
0.450 to 0.500 cm
-1
and
0.800 to 1.200 cm
-1
The currently valid cell con-
stant is displayed in the status
line.
LR325/001
LR 325/001 measuring cell,
nominal cell constant
0.010 cm
-1
.
The cell constant is perma-
nently set.
LR325/01
LR 325/01 measuring cell,
nominal cell constant
0.100 cm
-1
.
The cell constant can be ad-
justed in the range from 0.090
to 0.110 cm
-1
.
man
Any measuring cells with free-
ly adjustable cell constants in
the range from 0.250 to
25.000 cm
-1
.
Cell constant
0.090 to
0.110 cm
-1
Display and setting option of
the cell constant of the
LR 325/01 measuring cell.
Man. cell const.
0.250 to
25.000 cm
-1
Display and setting option of
the cell constant of any mea-
suring cells (man).
Menu item
Possible
setting
Description