Preparatory activities

Perform the following preparatory activities when you want to calibrate:

Carrying out an

ISE calibration

Proceed as follows to calibrate the instrument:

1

Connect the ISE combination electrode to the measuring in-
strument. The pH/mV/ISE measuring window is displayed.

2

Keep the standard solutions ready.

3

Measure the temperature of the standard solutions using a
thermometer.

1

In the measured value display, select the ISE measuring win-
dow with <

> <> and <M>.

2

Start the calibration with <CAL>.
The calibration display appears.

3

Thoroughly rinse the electrode with distilled water.

4

Immerse the electrode in standard solution 1.

5

Select the Set standard setting with <

> <> and press

<MENU/ENTER>.

6

Set the concentration of the standard solution with <

> <>

and press <MENU/ENTER>.

7

Measure the temperature of the standard solution using a ther-
mometer.

8

Select the Set temperature setting with <

> <> and press

<MENU/ENTER>.

9

Set the temperature with <

> <> and press

<MENU/ENTER>.

10

Select Continue with <

> <> and press <MENU/ENTER>.

The standard solution is measured.
The measured value is checked for stability (AutoRead).

i Immerse sensor in std. 1

Set temperature:

24 °C

Continue

Set standard:

0.010 mg/l

ISE <CAL>