Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 41

Multi 3500i
Operation
39
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the asymmetry of the elec-
trode.
8
Wait for the end of the AutoRead measurement or Terminate
AutoRead and accept the calibration value with
<MENU/ENTER>.
The calibration display for the next buffer appears.
9
For single-point calibration, select Exit with one point calibra-
tion with <
▲> <▼> and confirm with <MENU/ENTER>.
The calibration is completed as a single-point calibration.
The new calibration values are displayed as a message (
i
).
You have the following options:
z Accept the new calibration values with <MENU/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
z To switch to the measured value display without accepting
the new calibration values, press <M> or <ESC>.
i Buffer value = 6.80
i U = 12.0 mV
i Temperature =
24.8 °C
Terminate AutoRead
pH <CAL>
i Immerse sensor in buffer 2
i Temperature = 24.8 °C
Exit with one point calibration
Set buffer:
8.27
Continue
pH <CAL>