Multi 3500i

Operation

39

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the asymmetry of the elec-
trode.

8

Wait for the end of the AutoRead measurement or Terminate
AutoRead
and accept the calibration value with
<MENU/ENTER>.
The calibration display for the next buffer appears.

9

For single-point calibration, select Exit with one point calibra-
tion
with <

> <> and confirm with <MENU/ENTER>.

The calibration is completed as a single-point calibration.
The new calibration values are displayed as a message (

i

).

You have the following options:
z Accept the new calibration values with <MENU/ENTER>.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

z To switch to the measured value display without accepting

the new calibration values, press <M> or <ESC>.

i Buffer value = 6.80
i U = 12.0 mV
i Temperature =

24.8 °C

Terminate AutoRead

pH <CAL>

i Immerse sensor in buffer 2
i Temperature = 24.8 °C

Exit with one point calibration

Set buffer:

8.27

Continue

pH <CAL>