3 temperature compensation, Temperature compensation – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 62

Advertising
background image

Operation

Multi 3500i

60

4.7.3

Temperature compensation

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20 or Tr25.

You can select one of the following temperature compensation meth-
ods:

z Nonlinear temperature compensation (nLF)

according to EN 27 888

z Linear temperature compensation (Lin)

with selectable coefficients of 0.000 ... 3.000 %

/

K

z No temperature compensation (off)

Note
The reference temperature and temperature compensation are set in
the conductivity measuring menu (see section 4.7.4).

Application tips

Select the following temperature compensations given in the table ac-
cording to the respective test sample:

Test sample

Temperature compensation

Display
indicator

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous solu-
tions

lin
Set linear
temperature coefficient
0.001 ... 3.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT

Sal, nLF