2 measuring ranges, resolution, accuracy, 1 ph/orp, 2 ise – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 93: Measuring ranges, resolution, accuracy 7.2.1, Ph/orp

Advertising
background image

Multi 3500i

Technical data

91

7.2

Measuring ranges, resolution, accuracy

7.2.1

pH/ORP

7.2.2

ISE

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.000 ... + 20.000
- 2.00 ... + 20.00

0.001
0.01

U [mV]

- 999.9 ... + 999.9
- 2000 ... + 2000

0.1
1

T [°C]

- 5.0 ... + 105.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 20 ... + 130

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH *

± 0.004

+ 15 °C ... + 35 °C

U [mV] / range

- 999.9 ... + 999.9

± 0.2

+ 15 °C ... + 35 °C

- 2000 ... + 2000

± 1

+ 15 °C ... + 35 °C

T [°C] / temperature sensor

NTC 30

± 0.2

0 °C ... + 55 °C

PT 1000

± 0.3

0 °C ... + 55 °C

* when measuring in a range of ± 2 pH around a calibration point

Measuring ranges,

resolution

Variable

Measuring range

Resolution

ISE [mg/l]

0.000 ... 10.000
0.00 ... 100.00
0.0 ... 100.0
0 ... 2000

0.001
0.01
0.1
1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 20 ... + 130

1

Advertising