4 conductivity, Conductivity – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 95

Multi 3500i
Technical data
93
7.2.4
Conductivity
Measuring ranges,
resolution
Variable
Measuring range
Resolution
[µS/cm]
0.000 ... 2.000*
0.00 ... 20.00**
0.0 ... 200.0
0 ... 2000
0.001
0.01
0.1
1
[mS/cm]
0.00 ... 20.00
0.0 ... 200.0
0 ... 2000
0.01
0.1
1
Specific resistance
[kOhm*cm]
0.000 ... 2.000
0.00 ... 20.00
0.0 ... 200.0
0 ... 2000
0.001
0.01
0.1
1
Specific resistance
[MOhm*cm]
0.00 ... 20.00
0.0 ... 200.0
0 ... 2000
0.01
0.1
1
SAL
0.0 ... 70.0
according to the IOT
table
0.1
TDS [mg/l]
0 ... 2000
Factor can be set be-
tween 0.40 and 1.00
1
T [°C]
- 5.0 ... + 105.0
0.1
* only possible with cells of the cell constant, 0.010 cm
-1
** only possible with cells of the cell constant, 0.010 cm
-1
or 0.090 ... 0.110 cm
-1
Cell constants
Cell constant C
Values
Can be calibrated in
the ranges
0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
-1
Adjustable
0.010 cm
-1
(fixed)
0.090 ... 0.110 cm
-1
0.250 ... 25.000 cm
-1
Reference temperature
Reference
temperature
Values
Adjustable
20 °C (Tr20)
25 °C (Tr25)