What to do if...

Multi 3500i

84

No stable measured

value

Obviously incorrect

measured values

Sensor symbol flashes

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if

necessary

Electrode + test sample:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

pH electrode:

– Not connected

– Connect electrode

– Cable broken

– Replace electrode

– pH electrode unsuitable

– Use suitable electrode

– Temperature difference

between buffer and test
sample too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system