What to do if...

Multi 3500i

86

6.3

DO measurement

DO sensor was not

recognized

Error message OFL

Error message, Error

(invalid calibration)

Error message, LEAK

Measured value flashes

for a long time

(No stable measured

value)

Measured value too low

Cause

Remedy

– Sensor not connected

– Connect sensor

– Cable broken

– Replace sensor

Cause

Remedy

– Measuring range exceeded

– not possible

Cause

Remedy

DO sensor:

– Electrolyte solution depleted

– Regenerate sensor

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate sensor

Cause

Remedy

DO sensor:

– Membrane damaged

– Exchange the membrane cap

– Membrane head screwed on

not tight enough

– Screw membrane head tight

Cause

Remedy

DO sensor:

– Membrane contaminated

– Clean membrane

Cause

Remedy

DO sensor:

– Insufficient flow

– Provide flow to the sensor