Multi 3500i

Operation

57

5

Press <MENU/ENTER>.
The AutoRead measurement to determine the DO concentrati-
on starts.

6

Wait for the end of the AutoRead measurement or accept the
measured value with <MENU/ENTER>.
The determined DO content is displayed as a message (

i

).

7

Press <MENU/ENTER>.

8

Using <

> <>, set the correction factor to adjust the dis-

played concentration value to the nominal value (value of the
comparison measurement). Subsequently, accept the correc-
tion factor with <MENU/ENTER>.

9

Select Accept with <

> <> and press <MENU/ENTER>.

The measuring instrument switches to the measured value dis-
play. The [Factor] indicator appears in the measuring window.

i Concentration = 8.5 mg/l
i Temperature =

24.8 °C

Terminate AutoRead

ConOx <CAL>

i Concentration = 8.49 mg/l
i Temperature =

24.8 °C

Set factor:

1.000

Accept

ConOx <CAL>