Ac voltage measurements – B&K Precision 2516 - Manual User Manual

Page 112

Advertising
background image

112

WARNING:
Never connect more than 1000V DC across the

input terminals.


AC Voltage Measurements


AC voltage measurements use the same setup as DC voltage
measurements.

Configure all settings from the menu.

When set for Manual ranging (select Manual by toggling

), use the

key to increase the range. At the highest range, pressing this will

automatically go back to the lowest range.

AC Voltage ranges: 60 mV, 600 mV, 6 V, 60 V, 600 V, 750 V.

NOTE: When the measurement unit is changed, range

will automatically change to Auto ranging.

CAUTION:
Always connect the test leads to the instrument

inputs first before probing the DUT to avoid
potential shock hazard.


Follow these steps to make a measurement.

1.

Connect the black test lead to the COM input.

2.

Connect the red test lead to the V.Ω.C input.

3.

Probe with the test leads to the DUT and take the measured reading

on display.

NOTE: If the measurement is out of range or if manual

range is used and the measurement is outside of
the selected range, the instrument will beep
continuously.

Advertising
This manual is related to the following products: