Perform self-calibration, Figure 10 - boot screen, 4 perform self-calibration – B&K Precision 2516 - Manual User Manual

Page 31

Advertising
background image

31

The oscilloscope will display a boot screen shown below for 5-10 seconds
before the main operating display appears.

Figure 10 - Boot Screen

To power OFF the device, hold down

for at least 3 seconds.


2.4 Perform Self-Calibration


We recommend you to perform self-calibration if any of the following
applies:

-

Operating temperature changes by more than 5 °C after

30 minutes of use.

-

Abnormal behaviors and operations are present.

-

Measurements are out of accuracy.

CAUTION:

Before performing self-calibration, disconnect all
inputs of the instrument. Otherwise, the instrument
may not calibrate correctly or may cause errors.