Display menu – B&K Precision 2516 - Manual User Manual

Page 45

Advertising
background image

45

Acquisition

Sampling

Sample and display most waveforms accurately.

Peak Detect Detect burr and reduce fake wave phenomena.

Average

Reduce random and irrelative noise.

Averages

4, 16, 32,

64,128,256

Select the number of samples for average
acquisition.

Sinx/x

sinx

Use sine interpolation

x

Use linear interpolation

Mode

Equ time

Real time

Set the sampling mode to equivalent sampling.
Set the sampling mode to real time sampling.

Sa Rate

Display the sampling rate.


To change the settings to any of the menu options, press their
corresponding function keys (

-

).

Sampling: To construct the waveform, the scope samples the signals in
equal interval.

Peak Detect: The scope captures the maximum and minimum values of the
signals in every interval to display the waveform.

Average: The scope acquires several waveforms, averages them, and
displays the final waveform. The more average times the smoother the
waveform.

Equivalent Time Sampling: This mode is good for observing repetitive
periodic waveforms. The sampling rate is up to 50 GSa/s.

Real Time Sampling: The scope will use the highest real-time sampling rate
up to the specified maximum sampling rate.

Note: Sampling rate varies based on the selected horizontal timebase
setting.

Display Menu


To access the display menu, go to the scope menu and press

to

select Display. Below is a table of the Display menu structure.