Save recorded data to external memory – B&K Precision 2516 - Manual User Manual

Page 140

Advertising
background image

140

Record Data by Manual Trigger


By default, data is recorded automatically and continuously. To record data
by a manual trigger, go to the second page of the Meter Trend Plot menu

and press

to set the Manual option On or Off.

When set to On, the plot will stop recording. Instead, the

button

will function as a manual trigger button. Each subsequent press will trigger
one measurement recording to the plot.

NOTE:

In this mode, the

button will be used as a

trigger instead of entering the Recorder menu. To

exit out of Meter Trend Plot, press

to

select Return in the second page of Meter Trend
Plot menu. All previously recorded data will be
lost if you exit. A confirmation message will ask if
you want to exit.


Save Recorded Data to External Memory


The recorded data on the meter trend plot is stored temporarily. All
recorded data will be lost if the recording gets reset, exit out of the meter
trend plot mode, or power OFF the instrument.

However, the data can be saved onto an external USB flash drive.

To save:

1.

Go to the second page of the Meter Trend Plot menu and select CSV

by pressing

.

2.

The Directory menu will be displayed.

Advertising
This manual is related to the following products: