Note: Different timebase settings will result in different sampling rates,
both with normal and long memory depth configured in MemDepth.
Selecting a timebase lower than 50 ms will automatically set the
instrument into scan mode. Deep memory is not supported in scan mode.
For timebase selections faster than 25 µs, the deep memory function will
have an adverse affect and will result in fewer sample points than if deep
memory is disabled.

Below is a table describing the different sampling rates for different
horizontal timebase settings with and without deep memory enabled.

Table 13 - Sampling Rate Table

Timebase

Single channel

Interleaved

Normal
Sampling
Rate

Deep
Memory
Sampling
Rate

Normal
Sampling
Rate

Deep Memory
Sampling Rate

50 s

50 Sa/s

-

50 Sa/s

-

25 s

100 Sa/s

-

100 Sa/s

-

10 s

250 Sa/s

-

250 Sa/s

-

5 s

500 Sa/s

-

500 Sa/s

-

2.5 s

1 kSa/s

-

1 kSa/s

-

1 s

2.5 kSa/s

-

2.5 kSa/s

-

500 ms

5 kSa/s

-

5 kSa/s

-

250 ms

10 kSa/s

-

10 kSa/s

-

100 ms

25 kSa/s

-

25 kSa/s

-

50 ms

25 kSa/s

1 MSa/s

25 kSa/s

1 MSa/s

25 ms

50 kSa/s

2.5 MSa/s

50 kSa/s

2.5 MSa/s

10 ms

125 kSa/s

5 MSa/s

125 kSa/s

5 MSa/s

5 ms

250 kSa/s

10 MSa/s

250 kSa/s

10 MSa/s

2.5 ms

500 kSa/s

25 MSa/s

500 kSa/s

25 MSa/s

1 ms

1.25 MSa/s

50 MSa/s

1.25 MSa/s

50 MSa/s

500 µs

2.5 MSa/s

100 MSa/s

2.5 MSa/s

100 MSa/s

250 µs

5 MSa/s

5 MSa/s

100 µs

12.5 MSa/s

12.5 MSa/s

50 µs

25 MSa/s

25 MSa/s

25 µs

50 MSa/s

50 MSa/s

10 µs

100 MSa/s

100 MSa/s