B&K Precision 2516 - Manual User Manual

Page 147

Advertising
background image

147

2 mV/div: ≤ ± 4 %

Max. BNC Input
Voltage

CATII 300 V RMS from BNC signal to BNC shell

Max. input
Voltage for
probe

(4)

1x/10x CATII 300 V RMS (PR150SA)

(5)(6)

10x CATIII 600 V RMS, CATII 1000 V RMS (PR250SA)

Channel Voltage
Offset Range

2mV - 200mV : ±1.6V
206mV - 10V : ±40V
10.2V - 100V : ±400V

5mV - 200mV : ±1.6V
206mV - 10V : ±40V
10.2V - 100V : ±400V

Bandwidth Limit 20MHz (-3dB)
Horizontal Scan
Range

5.0nS/div -
50S/div

2.5nS/div -

50S/div

5.0nS/div -
50S/div

2.5nS/div -

50S/div

Timebase
Accuracy

± 50 ppm measured over 1 ms interval

Input Coupling

AC, DC, GND

Input
Impedance

1 MΩ +/- 2 % || 18 pF ± 3
pF

1 MΩ +/- 2 % || 16 pF ± 3
pF

Probe
Attenuation
Selectable
Factors

1X, 5X, 10X, 50X, 100X, 500X, 1000X

Vertical and
Horizontal
Zoom

Vertically or horizontally expand or compress a live or
stopped waveform.

I/O Interface

USB

USB host port support USB flash drives (FAT format).
USB device port for PC connectivity

Acquisition Modes
Sampling

Display sample data only

Peak Detect

Capture the maximum and minimum values of a signal

Average

Waveform averaged, selectable from 4, 16, 32, 64, 128,
256

Trigger System

Trigger Types

Edge, Pulse Width, Video*, Slope, Alternative

*Support signal Formats: PAL/SECAM, NTSC
Trigger condition: odd field, even field, all lines, or line
number