Table 16 - Edge Trigger Setup Menu

Menu

Option

Setting

Description

Coupling

DC

Passes all components of the signal

AC

Blocks DC components and attenuates
signals below 170 Hz.

HF Reject

Attenuates the high-frequency
components above 140 kHz.

LF Reject

Blocks the DC component and attenuates
the low-frequency components below 7
kHz.

Hold off

Adjusts the hold off time.

Hold off

Reset

Reset hold off time to 100 ns.

Return

Return the Edge Trigger Menu.

Hold Off: To adjust the hold off time, go to the Edge Trigger Setup Menu

and press

, then use the

or

arrow keys to adjust

the hold off time, which will be displayed when it is adjusted.

Hold off time can be reset to 100 ns by pressing

to select Hold Off

Reset.