8 Specifications


Oscilloscope Specifications


All specifications apply to measurements with a 10X probe. To verify that
the oscilloscope meets specifications, the oscilloscope must first meet the
following conditions:

The oscilloscope must have been operating continuously for 30

minutes within the specified operating temperature.

If the operating temperature changes by more than 5° C, you must

perform the Do Self Cal operation, accessible through the Utility
menu.

All specifications are guaranteed unless noted “typical.”

Models

2511

2512

2515

2516

Bandwidth

60 MHz

100 MHz

60 MHz

100 MHz

Real Time
Sampling Rate

1 GSa/s (half-channel interleaved)

(1)(2)

, 500 MSa/s (per

channel)

Channels

2 (non-isolated)

2 (isolated)

Rise Time

< 5.8 ns

< 3.5 ns

< 5.8 ns

< 3.5 ns

Ch to Ch
Isolation (Both
channels in
same V/div
setting)

> 100:1 at 50 MHz

Memory Depth

40 kpts (half-channel interleaved)

(1)(2)

, 20 kpts (per

channel)

Deep Memory

(3)

2 Mpts (half-channel interleaved)

(1)

, 1 Mpts (per channel)

Vertical
Resolution

8 bits

Vertical
Sensitivity

2 mV/div – 100 V/div (1-2-5
order)

5 mV/div – 100 V/div (1-2-5
order)

DC Gain
Accuracy

5 mV/div-100 V/div: ≤ ± 3
%

5 mV/div-100 V/div: ≤ ± 3
%