Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 20
WaveJet 300A Series
20
917976 Rev A
Hard Copy
COPY
Command
DESCRIPTION
The COPY command copies the current screen data to USB
Memory.
COMMAND SYNTAX
COPY
QUERY SYNTAX
RESPONSE FORMAT
Vertical CPL
Command/Query
DESCRIPTION
The CPL command sets the coupling mode of the specified
input channel. The CPL? query returns the selected
coupling mode.
COMMAND SYNTAX
<channel>:CPL <coupling>
<channel>:={C1, C2, C3, C4} (for WJ354/334/324/314)
:={C1, C2} (for WJ352/332/322/312)
<coupling>:={AC1M, GND, DC1M, DC50} (for
WJ354/334/352/332)
:={AC1M, GND, DC1M} (for
WJ324/314/322/312)
QUERY SYNTAX
<channel>:CPL?
RESPONSE FORMAT
<coupling>
The current coupling mode is identified in <coupling>.
Cursor CURM
Command/Query
DESCRIPTION
The CURM command defines the cursor mode. The CURM?
query returns the defined cursor mode.
COMMAND SYNTAX
CURM <mode>
<mode>: = {OFF, DV, DH, DHDV, VATH}
QUERY SYNTAX
CURM?
RESPONSE FORMAT
<mode>
System DATE
Command/Query
DESCRIPTION
The DATE command defines the date and time. The DATE?
query reads the defined date and time.
COMMAND SYNTAX
DATE <day>,<month>,<year>,<hour>,<minute>,<second>
<day>:= a number from 1 to 31.
<month>:= {JAN, FEB, MAR, APR, MAY, JUN, JUL, AUG, SEP,
OCT, NOV, DEC}
<year>:= a number from 2000 to 2099.
<hour>:= a number from 0 to 23.
<minute>:= a number from 0 to 59.
<second>:= a number from 0 to 59.
QUERY SYNTAX
DATE?
RESPONSE FORMAT
<day>,<month>,<year>,<hour>,<minute>,<second>