Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 43
WaveJet 300A Remote Control Manual
917976 Rev A
43
Horizontal TDIV
Command/Query
DESCRIPTION The
TDIV command sets timebase. The timebase is specified
using any of the following suffixes: NS (nanosecond), US
(microsecond), MS (millisecond), or S (second). If the
specified value exceeds the allowable range, it is
rounded to a permissible value. The TDIV? query returns
the currently set timebase.
COMMAND SYNTAX
TDIV <value>
<value>:= Timebase setup parameter. The <DECIMAL NUMERIC
PROGRAM DATA> format is used, and the suffix is also
valid.
QUERY SYNTAX
TDIV?
RESPONSE FORMAT
<value>
<value>:= The currently set timebase is identified in
<NR3 NUMERIC RESPONSE DATA> format.
REMARKS
The TDIV command rounds the figure in <value> as follows.
Rounding range
Set timebase
Note
value <= 500E-12
500 ps/div
500 ps/div is available for
WJ354/352 only
500E-12 < value <= 1E-9
1 ns/div
1 ns/div is available for
WJ354/352/334/332
1E-9 < value <= 2E-9
2 ns/div
2 ns/div is available for
WJ354/352/334/332/324/322
2E-9 < value <= 5E-9
5 ns/div
5 ns/div is available for all models
: :
10 < value <= 20
20 s/div
20 < value
50 s/div