Automatic Measurement

MSEL

Command/Query

DESCRIPTION

The MSEL command selects the measurement item of
automatic measurement A, B, C or D defined using the
automatic measurement direction command DIRM. The
MSEL? query returns the selected measurement item.

COMMAND SYNTAX

MSEL <ch>, <mode>
<ch>:={OFF, CH1, CH2, CH3, CH4, MATH}
(WJ354/334/324/314)
:={OFF, CH1, CH2, MATH} (WJ352/332/322/312)
<mode>:= {MAX, MIN, P-P, VRMS, CVRMS, VMEAN, CVMEAN,
TOP, BASE, T-B, +OSHOT, -OSHOT, TR20-80, TF80-20,
TR10-90, TF 90-10, FREQ, PERIOD, +PULSE, -PULSE,
+WIDTH, -WIDTH, DUTY, INTEGRAL, SKEW, DELTAT}

QUERY SYNTAX

MSEL?

RESPONSE FORMAT

<ch>, <mode>
The selected automatic measurement channel is
identified in <ch>.
The automatic measurement item is identified in <mode>.

REMARKS

If <ch> is set to CH3 or CH4 in the model
WJ352/332/322/312, it is rounded to CH1.

Automatic Measurement

MSRA?, MSRB?, MSRC?, MSRD?

Query

DESCRIPTION MSRA?,

MSRB?, MSRC? and MSRD? queries respectively read

automatic measurements A, B, C, and D.

COMMAND SYNTAX

QUERY SYNTAX

MSRA? (or MSRB?, MSRC?, or MSRD?)

RESPONSE FORMAT

<measure>
The result of automatic measurement A (or B, C, or D)
is identified in <measure>. The indicated value is
converted into <NR3 Numeric Response Data> and
returned. If automatic measurement is not possible,
+9.91E+37 is returned.

<value>
<value>, <maximum_value>, <minimum_value>
If automatic measurement is not valid (in case
measurement is turned off), the result becomes
+9.91000E+37.