Math FHZPOS

Command/Query

DESCRIPTION

The FHZPOS command sets the frequency at the center of
the screen in Hz, to set horizontal position of FFT
waveform.
The FHZPOS? query returns the frequency at the center
of the screen of FFT waveform in Hz.

COMMAND SYNTAX

FHZPOS <fft_hpos>
<fft_hpos> is set in <DECIMAL NUMERIC PROGRAM DATA>
format. A value beyond the setting range is rounded to
a permissible closer value.

QUERY SYNTAX

FHZPOS?

RESPONSE FORMAT

<fft_hpos>
<fft_hpos>:= <NR3 Numeric Response Data> format.
<fft_hpos> shows the horizontal position (position) of
the FFT waveform.

REMARKS

Math FHZZOOM

Command/Query

DESCRIPTION

The FHZZOOM command sets the horizontal zoom factor of
the FFT waveform.
The FHZZOOM? query returns the horizontal zoom factor
of the FFT waveform.

COMMAND SYNTAX

FHZZOOM <fft_hzoom>
<fft_hzoom>:= {1, 2, 5, 10}
A value other than previously specified is rounded to
a permissible larger value.

QUERY SYNTAX

FHZZOOM?

RESPONSE FORMAT

<fft_hzoom>
<fft_hzoom> shows the horizontal zoom factor of the FFT
waveform.

REMARKS

When FFT trace is off, FHZZOOM command reports

command-execution error (CME).

Horizontal FRQCNT?

Query

DESCRIPTION

The FRQCNT? query returns the trigger signal frequency
result that is measured by internal trigger frequency
counter.

COMMAND SYNTAX

QUERY SYNTAX

FRQCNT?

RESPONSE FORMAT

<value>
<value>:= <NR3 Numeric Response Data> format. <value>
shows the frequency of trigger signal.