1 wj-a-gpib hardware, 2 wj-a-lan hardware, Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual

Page 5

Advertising
background image

WaveJet 300A Remote Control Manual

1. INTRODUCTION TO WAVEJET 300A REMOTE CONTROL INTERFACES

The WaveJet 300A supports remote control through USB, GPIB and TCP/IP. USB is supported as a standard
feature, GPIB and TCP/IP are available with optional interface boards.

The WJ-A-GPIB and WJ-A-LAN are optional interface boards exclusively for the WaveJe 300A series.

• Insert the board into one of the option slots on the rear of the WaveJet 300A before use.
• It can be inserted into either of the two option slots provided.
• Once the board is inserted into the slot, the WaveJet main unit automatically recognizes it as a remote

interface, allowing you to select it.

• The WJ-A-GPIB and the WJ-A-LAN can be inserted at the same time, but only one of them can be selected.
• Two of the same kind of interface board must not be inserted at the same time (prohibited).

CAUTION!

To prevent damage to the instrument, turn instrument power OFF before inserting a WJ-A-GPIB or
WJ-A-LAN interface board.

ESD Sensitive

The WJ-A-GPIB and WJ-A-LAN interface boards are sensitive to electrostatic discharge
(ESD). To avoid causing damage to the boards, always follow anti-static procedures (wear
wrist strap, etc.) when handling the boards.

1.1 WJ-A-GPIB Hardware

Applicable standard

GPIB IEEE-488.1

Usable unit

WaveJet 3xx series firmware 3.xx or later

Power supply

Supplied from WaveJet3xx option slot

Power consumption

Included in WaveJet3xx main unit

Dimensions

63 mm

× 80 mm (excluding protrusion of connector terminal)

Mass Approximately

70

g

Operating temperature range

0 to 40

°C

Operating humidity range

80% RH or lower (when operated within 0 to 40

°C)

Storage temperature range

-20 to

+60 °C

1.2 WJ-A-LAN Hardware

Applicable standard

TCP/IP 10BASE-T/100BASE-T

Usable unit

WaveJet 3xx series firmware 3.xx or later

Power supply

Supplied from WaveJet3xx option slot

Power consumption

Included in WaveJet3xx main unit

Dimensions

63 mm

× 80 mm (excluding protrusion of connector terminal)

Mass

Approximately 55 g

Operating temperature range

0 to 40

°C

Operating humidity range

80% RH or lower (when operated within 0 to 40

°C)

Storage temperature range

-20 to

+60 °C

917976 Rev A

5

Advertising