Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual

Page 55

Advertising
background image

WaveJet 300A Remote Control Manual

917976 Rev A

55

Cursor VCUR

Command/Query

DESCRIPTION

The VCUR command defines the position of the voltage
cursor. The VCUR? query returns the defined position.

COMMAND SYNTAX

VCUR <Cursor1>,<Cursor2>
<Cursor1>:=-3.97 to +3.95
<Cursor2>:=-3.97 to +3.95
<Cursor1> and <Cursor2> respectively represent
positions of voltage cursors Cursor1 and Cursor2 on the
screen, using the unit "div". For information on how
to round the value, see "Remarks". The bottom end of
the screen corresponds to -3.97, the center to 0, and
the top end to +3.95.

QUERY SYNTAX

VCUR?

RESPONSE FORMAT

<Cursor1>, <Cursor2>
Current positions of the voltage cursors are identified
in <Cursor1> and <Cursor2>.
The format is <NR3 Numeric Response Data>.

REMARKS

The resolution of <Cursor1> and <Cursor2> is 0.02. If
other value is input, it is rounded to the number having
small absolute value.
VCUR command reports Device-specific error (DDE) when
CURM is not DV or DHDV.

Advertising