Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual

Page 49

Advertising
background image

WaveJet 300A Remote Control Manual

917976 Rev A

49

(Delay value resolution) = (Sweep time) / 50
• When the specified resolution is not equal to the above value, it is rounded
to the permissible value of which the absolute value is smaller.

Acquisition *TRG

Command

DESCRIPTION

The *TRG command sets the trigger mode to "SINGLE" to
start single waveform acquisition.

COMMAND SYNTAX

*TRG

QUERY SYNTAX

RESPONSE FORMAT

REMARKS

*TRG reports device-specific error (DDE) and is not
performed when XYDS is XY.

Acquisition TRMD

Command/Query

DESCRIPTION

The TRMD command selects the trigger mode. The TRMD?
query returns the currently selected trigger mode.

COMMAND SYNTAX

TRMD <mode>
<mode> := {AUTO, NORMAL, SINGLE, STOP}

AUTO := Set trigger mode to AUTO when scope is not in
roll mode operation. If roll mode is on and scope is
in roll mode operation range, AUTO initiates ENDLESS
roll acquisition.
NORMAL := Set trigger mode to NORMAL when scope is not
in roll mode operation. If roll mode is on and scope
is in roll mode operation range, NORMAL initiates
repeated TRIG'D roll acquisitions.
SINGLE := Set trigger mode to NORMAL when scope is not
in roll mode operation. If roll mode is on and scope
is in roll mode operation range, SINGLE initiates
single TRIG'D roll acquisition.

QUERY SYNTAX

TRMD?

RESPONSE FORMAT

<mode>
Not in roll mode operation: AUTO, NORMAL, SINGLE or STOP
In roll mode operation: ENDLESS, TRIG'D, TRIG'D or STOP

REMARKS

TRMD SINGLE reports device-specific error (DDE) and is
not performed when XYDS is XY.