Trigger TVSET
Command/Query
DESCRIPTION The
TVSET command sets field sequence, field number and
line number for TV trigger.
The TVSET query returns the currently set field
sequence, field number, and line number.
COMMAND SYNTAX
TVSET <seq>,<field>,<line>
<seq> := {1, 2, 4, 8} This specifies the field sequence
to 1, 2, 4 or 8.
<field> := 0, 1 to 8 (Specify 0 to selects "Any Fields")
<line> := 1 to 263 (When TVSTD is NTSC and <field> is
odd number)
:= 1 to 262 (When TVSTD is NTSC and <field>
is even number)
:= 1 to 313 (When TVSTD is PAL and <field> is
odd number)
:= 314 to 625 (When TVSTD is PAL and <field>
is even number)
:= 0 (Specify 0 to selects "Any Lines")
QUERY SYNTAX
TVSET?
RESPONSE FORMAT
<seq>,<field>,<line>
REMARKS
TVSET command reports device-specific error (DDE) when
TTYP is not TV.
Line number selection is automatically set as follows.
NTSC:263/262
PAL:625
Trigger TVSTD
Command/Query
DESCRIPTION The
TVSTD command selects the TV standard for TV trigger
(TTYP=TV).
The TVSTD? query returns the currently selected TV
standard.
COMMAND SYNTAX
TVSTD <type>
<type> := {NTSC, PAL, CUSTOM}
QUERY SYNTAX
TVSTD?
RESPONSE FORMAT
<type>
REMARKS
TVSTD command reports device-specific error (DDE) when
TTYP is not TV.