917976 Rev A

Status TESE

Command/Query

DESCRIPTION The

TESE

command

sets the trigger event status enable

register. In other words, it determines whether the
summary message of the trigger event status register
is connected to the status byte. The TESE? query reads
the current setting of the trigger event status enable
register.

COMMAND SYNTAX

TESE <value>
<value>:=0 to 255

QUERY SYNTAX

TESE?

RESPONSE FORMAT

<value>
The current setting of the trigger event status enable
register is identified in <value>.
The format is <NR1 Numeric Response Data>.

Status TESR?

Query

DESCRIPTION

The TESR? query reads the trigger event status
register. In other words, it determines whether a
single measurement is completed. After the status is
read, the contents of the register are cleared.

COMMAND SYNTAX

QUERY SYNTAX

TESR?

RESPONSE FORMAT

<status>
Each bit of the trigger event status register is
weighted by a power of 2. The format is <NR1 Numeric
Response Data> and the value ranges from 0 to 255.

Trigger THTM

Command/Query

DESCRIPTION The

THTM command defines the trigger hold-off time. The

THTM? query returns the defined trigger hold-off time.

COMMAND SYNTAX

THTM <holdoff>
<holdoff>:= The <DECIMAL NUMERIC PROGRAM DATA> format
is used, and the suffix is also valid.
Available range is 0 to 50.0 s.
If specified holdoff is < 200 ns, the trigger hold-off
is turned off.

QUERY SYNTAX

THTM?

RESPONSE FORMAT

<holdoff>
The currently set hold-off time is identified in
<holdoff>. If the hold-off time is set to
"OFF", 0 is input to <holdoff>. The format is <NR3
NUMERIC RESPONSE DATA>.

REMARKS

THTM command reports device-specific error (DDE) when
TTYP is not EDGE..