Teledyne LeCroy WaveJet 300A Series Remote Control User Manual

Page 52

Advertising
background image

WaveJet 300A Series

52

917976 Rev A

Trigger TTYP

Command/Query

DESCRIPTION

The TTYP command selects the trigger type.
The TTYP? query returns the currently selected trigger
type.

COMMAND SYNTAX

TTYP <type>
<type> := {EDGE, PULSE, PERIOD, COUNT, TV}

QUERY SYNTAX

TTYP?

RESPONSE FORMAT

<type>
<type>:= Set trigger type

REMARKS

Trigger TVCUST

Command/Query

DESCRIPTION

The TVCUST command defines field frequency and number
of scan line for custom TV trigger.
The TVCUST query returns the currently defined field
frequency and number of scan line.

COMMAND SYNTAX

TVCUST <field>,<line>
<field> := 10 to 99. Then the field frequency is set
to 1/<field>.
<line> := 100 to 3000. Then the number of scan lines
is set to <line>.

QUERY SYNTAX

TVCUST?

RESPONSE FORMAT

<field>,<line>

REMARKS

TVSTD command reports device-specific error (DDE) when
TTYP is not TV or TVSTD is not CUSTOM.

Advertising