Status *SRE

Command/Query

DESCRIPTION

The *SRE command defines the service request enable
(SRE) register. This command allows the user to specify
what SBT register summary bit is used to create the
service request (SRQ). If "1" is written to the
associated bit location, the summary message bit is
enabled. If "0" is written to the associated bit
location, the service request is no longer created by
the associated event. After the SRE register is cleared,
the SRQ interrupt is disabled. When the value is
converted into binary code, the *SRE? query
returns the value that represents the bit setting in the
SRE register.

COMMAND SYNTAX

*SRE <value>
<value>:=0 to 255

QUERY SYNTAX

*SRE?

RESPONSE FORMAT

<value>
The current SRE register settings are identified in
<value> in <NR1 Numeric Response Data> format.

Status *STB?

Query

DESCRIPTION The

*STB? query reads the status register (STB) defined

in IEEE 488.1 and the master summary status (MSS). The
response includes those values specified in bits 0 to
7 of the status byte register.
For details about structure of each bit, see Remarks.
The response to the *STB? query is the same as that for
series polling, except that bit 6 contains the MSS
summary message instead of the RQS message.

COMMAND SYNTAX

QUERY SYNTAX

*STB?

RESPONSE FORMAT

<value>
The current STB register settings are identified in
<value> in <NR1 Numeric Response Data> format.

Acquisition STOP

Command

DESCRIPTION STOP

sets the trigger mode to "STOP". If the trigger mode

has already been "STOP", this command is ignored. STOP
acts same as "TRMD STOP".

COMMAND SYNTAX

STOP

QUERY SYNTAX

RESPONSE FORMAT