7.8 Temperature functions (Basic or detailed Setup)

Temperature

To set the functions of temperature measurement.
Notice: Standard range: -200 to 230°C
High temp. range (/HT): 0 to 400°C

Temperature Unit

Select:

degC, degF, K

Temperature format

Select:

xxxxxxx, xxxxx.X, xxxx.XX, xxx.XXX, xx.XXXX, x.XXXXX

Temperature lrv
(low range value)

Set a temperature value to the 4 mA or 0 Hz

Temperature Urv
(upper range value)

Set a temperature value to the 20 mA or xxxxx Hz.
Abs(URV-LRV) ≥ 10K

Temperature damping

Set the

damping time in s. Damping time affects outputs and display.

After 5 times damping time the given value is displayed (t=63%).

Temperature alm 1 sel

Select:

No function, More than, less than

Temperature alm 1 crit

Select a

temperature alarm level.

What happens in case of an alarm has to be defined in Status out (7.12,
7.13).

Temperature alm 2 sel

Select:

No function, More than, less than

Temperature alm 2 crit
Restrictions:
All alarms have to be in 0-110%
of URV-LRV Alarm criteria 2
must always be larger than alarm
criteria 1.

Select a

temperature alarm level.

What happens in case of an alarm has to be defined in Status out (7.12,
7.13).

Temperature fix val select

Select:

Inhibit, Enable

Temperature fixed value

Select a temperature value in case of malfunction of the built in
temperature sensor.

Temperature gain

The ROTAMASS temperature sensor is installed outside of the measuring
tubes. So at high or low fluid temperature the temperature measurement
may show a deviation as a function of the quality of heat insulation. This
function allows correcting temperature measurement for better accuracy
in mass flow and density.
Note: It is not recommended to use this function unless correct
measurement over the complete process temperature range can be
confirmed. It is not sufficient to proof correct measurement at one
temperature or in a very narrow temperature range unless the process
operates exclusively within this range.