8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-10

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.7 disagreement of indication with actual temperature

yes

Are temperature unit,

URV, LRV and lowcut set

correctly ?

Set correct temperature

parameters.

PT 100 fault.

Simulate the correct

temp. by a fixed temp,

value or connect an

external PT100.

For Remote type: Is

remote cable connected

according chapter 9 and

4.8.4 ?

Connect remote cable

according chapter 9

and 4.8.4

Compact type: disconnect first the converter housing

Disconnect the temperature sensor cables 7,8,9 from TP1, TP2, TP3 in

detector RCCS3x terminal box.

Measure the resistances of detector temperature sensor terminals TP1,

TP2, TP3.
Resistance between TP1 + TP2 : 50 …200

depending on medium

temperature
Resistance between TP1 + TP3 : 50 …200

depending on medium

temperature
Resistance between TP2 + TP3 : <10

Replace remote cable.

Is fixed temperature

selected (Tempy fix val sel

= Enable ) ?

Set Temp fix val sel =

Inhibit.

Resistance values are as

stored ?

For remote type: Check

remote cable lines 7, 8, 9.

Cable OK ?

yes

n o

no

no

no

no

yes

yes

yes

yes

F87.EPS