4.8.5 power supply wiring

The terminals for the power supply and for the I/O ports are in a separate terminal box as shown in the
figures below.

Cable connection to rack-mount converter RCCR31:
1.

Loose the cable clamp on terminal board

2.

Remove the 25 mm section of PVC outer sheathing from the cable and fix the outer ring of the cable

with the cable clamp on terminal board. (see left figure below).

3.

Connect the numbered leads to the terminals as shown in the right figure below.

4.

Connect inner shields to terminal COM.

Detector

Cable

Converter

RCCS3

RCCY03

RCCR31

D+

1

1

D+

D-

2

2

D-

S1+

3

3

S1+

S1-

4

4

S1-

S2+

5

5

S2+

S2-

6

6

S2-

TP1

7

7

TP1

TP2

8

8

TP2

TP3

9

9

TP3

Shields

COM