8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-8

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.5 disagreement of indication with actual flow rate

no

yes

no

yes

no

yes

no

no

no

Run autozero according

chapter 5.6.

Are mass flow unit, URV,

LRV and lowcut set

correctly ?

Set correct mass flow

parameters.

Contact the nearest

YOKOGAWA office or

service center.

Is electrical grounding

correct according chapter

4.8.4; 4.8.5; 9 ?

Make correct grounding.

Control the piping. Is

there a location where

air can be sucked in?

Connect remote cable

according chapter 9

and 4.8.1.

Is zero point set correctly?

Is gas penetration in the

fluid (chek the density)?

yes

Is the control

measurement correct and

reliable?

For Remote type: Is

remote cable connected

according chapter 9 and

4.8.1 ?

yes

yes

Check the control

measurement.

F85.EPS