8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-6

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.2 No key-setting possible

8.7.3 No HArT communication

no

yes

no

yes

Is parameter Key Status

set enable ?

Set parameter Key

Status enable .

Is display cable connected

correct to CPU board ?

Connect the cable in the

correct position.

Contact the nearest

YOKOGAWA office or

service center.

F82.EPS

no

yes

no

yes

Is HART DD of Rotamass

RCCT3 available in HART

communicator ?

Store necessary DD in

HART communicator.

Is HART communicator

connected to Rotamass

according chapter 4.8.7 ?

Connect HART

communicator according

chapter 4.8.7.

Contact the nearest

YOKOGAWA office or

service center.

F83.EPS