6 disagreement of indication with actual density – Yokogawa RotaMASS 3-Series User Manual

Page 185

Advertising
background image

8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-9

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.6 disagreement of indication with actual density

no

yes

yes

no

no

yes

no

yes

no

yes

yes

no

yes

no

Set correct density

parameters.

Change piping . Vertical

installation is

recommended.

Check installation.

Connect remote cable

according chapter 9

and 4.8.4.

Make troubleshooting

according chapter 8.7.7.

Set Density fix val sel =

Inhibit.

Is temperature

measurement correct ?

For Remote type: Is

remote cable connected

according chapter 9 and

4.8.4 ?

Let YOKOGAWA perform

a new density and mass

flow calibration.

Has corrosion or errosion

occured (too high

density reading)?

Contact the nearest

YOKOGAWA office or

service center.

Is fixed density selected

(Density fix val sel =

Enable ) ?

Is the tube filled with liquid

completely and without

bubbles (density low

reading)?

Is there sediment or
clogging in the tubes ?

Clean the tubes by means

like water, steam or other

solvent.

Are density unit, URV,

LRV and lowcut set

correctly ?

F86.EPS