6. OpErATION vIA HArT

6-5

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

6.3 Communication via HArT Communicator

Keys and functions of HHT 375:

Touch screen display

HART communication
terminals

Enter key

Function key

Alphanumeric keypad

On/Off key

Navigation keys
(4 arrow keys)

Tab key

Hot key

The HART communicator automatically searches for ROTAMASS RCCT_F3 on the 4-20mA loop, when it is
turned on. When HART communicator is connected to RCCT_F3, it displays the online menu.
(If RCCT_F3 is not found, it displays “No device found. Press OK…”. Press OK ´F4´ function key and the
main menu appears. Please retry after confirming the connection with RCCT_F3).
If there is no device specific DD installed in the HHT375 (RCCT_F3, Dev v4, DD v1), the functionality of the
HART Menu as listed in chapter 7.4 is not total supported.
In this case the HHT375 works in the “Generic Mode”, which allows configuring the ROTAMASS with limited
functionality.
For upgrading your HHT375 please use "Easy Upgrade" programming utility which is available from
EMERSON.
Online Menu Summary

No.

Display item

Contents

1

Device Setup

Set parameters for ROTAMASS

2

PV

Display process value in engineering unit

3

PV AO

Display analog output in mA

4

Review

Review parameters for ROTAMASS

5

Service

Not open