7. pArAMETEr dESCrIpTION

7-90

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

How to set status, pulse or frequency output
(using the display keys, equivalent via HART)
Status

Basic Setup or detailed Setup

pulse/Status out 1
pulse/Stat 1 select
Select :

Status out

pulse or Frequency

Basic Setup or detailed Setup

pulse/Status out 1

pulse/Stat 1 select
Select :

pulse

pulse 1 select

F-Total mass

pulse

r-Total mass

pulse

F-Total volume

pulse

r-Total volume

pulse

F-Total net

pulse

r-Total net

pulse

Mass Flow

Frequency

volume Flow

Frequency

Net Flow

Frequency

density

Frequency

Temperature

Frequency

Concentration meas.

Frequency

Notice : By selecting a variable (mass or volume)
which can be totalized automatically the pulse
output is selected.

SO 1 function

Select:

No Function, Flow direction, Total limit switch, Mass flow

alm 1, Mass flow alm 2, Mass flow alm 1+2, vol flow alm 1, vol flow
alm 2, vol flow alm 1+2, density alm 1,density alm 2, density alm
1+2, Temp alm 1, Temp alm 2, Temp alm 1+2, Concentration meas
alm 1, Concentration meas alm 2, Concentration meas alm 1+2,
Net flow alm 1, Net flow alm 2, Net flow alm 1+2, Slug alarm, Empty
alarm, Corrosion alarm, All alarms, All errors, All alarms & errors

If

Flow direction is selected, status output becomes active for forward

flow and not active for reverse flow, if parameter

flow direction set to

forward. It is the other way round, if parameter flow direction is set to
reverse (see 7.17).

If

Total limit switch is selected, status output becomes active, when

the totalizer reaches the value set to parameter

Total limit switch (see

7.17).

If an alarm function is selected, status output becomes active, if the
concerning alarm occurs.
If

All alarms is selected, status output becomes active, if at least one

alarm occurs. The same applies for errors with the setting All errors.
For

All alarms & errors the status output becomes active, if at least

one alarm or one error occurs.

SO1 active mode

Select:

On Active, Off Active

The output levels depending on this parameter are shown on page 7-91.

If

pulse is selected, under Basis Setup the

pulse unit and

pulse rate

have to be defined.
Under ´Detailed Setup´

pulse unit

pulse rate
pulse active mode
and

pulse width

have to be defined.
If the pulse output is set under

Basic Setup, the

pulse width is the width set under

detailed Setup

If the pulse width is not configured under

detailed

Setup menu the pulse width will be set default to
1ms.
If ´frequency´ is selected the frequency generated
at URV (upper range value) has to be defined.
Default is URV = 1000Hz.
differences between pulse and Frequency
output
For pulse output the mass (or volume) is internally
totalized and if the increment is reached, one
pulse is sent.
For a frequency output the measured value is
calculated in actual frequency according to the
definition of the user (e.g. 10000 Hz = 54321 kg/h)
with 50% duty cycle.
Set as frequency output the pulse length is changing
with the change of the measured value.