Mechanical specifications – Yokogawa RotaMASS 3-Series User Manual

Page 225

Advertising
background image

11. TECHNICAl dATA

11-3

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

Other 2 phase Flow, liquid/solid and liquid/liquid
Two phase flow can generate minus span errors. The errors
are proportional to the difference in density between the 2
phases and the amount of the second phase. If the particles
(or droplets) are very small no errors will be generated.

power supply and power Consumption
- AC-type

: 90 to 264 V AC, 47-63 Hz

For Ex version 250 V AC max.

- DC-type

: 20.5 to 28.8 V DC

Consumption

: max. 25VA / 10W

MECHANICAl SpECIFICATIONS

protection Class

- RCCT3x

: IP66/67

- RCCF31

: IP66/67

- RCCS3x

: IP66/67

- RCCS3x with /Tx

: IP54

- RCCR31

: IP20

Materials

- Detector housing

: Stainless steel 304/1.4301

- Detector terminal box : 316L/1.4404

- Detector gas filling plug: 1.4305

- Detector rupture disk (/RD) : 316L

- Field- mount converter housing

: Aluminium alloy with Polyurethane

corrosion-resistant coating or

epoxy coating (option /X1)

- Rack- mount converter housing

: Aluminium

Coating Color

- Field-mount converter case

: Mint green

Wetted parts

- RCCS30 to 33 :

Tubes

: Hastelloy C-22/2.4602

Process connections : 316L / 1.4404

- RCCx34 to 39/IR :

Tubes and process connection

: 316L / 1.4404/1.4435 or

Hastelloy C-22/2.4602

- RCCx39/XR :

Tubes and process connection

: 316L / 1.4404

Table 11-5 : Diameter of measuring tubes

Type

RCCS30 RCCS31 RCCS32 RCCS33

Inner
diameter

mm

1.2

2.1

3

4.5

Wall thick-
ness

mm

0.2

0.25

0.25

0.25

Type

RCCx34 RCCx36 RCCx38 RCCx39

RCCx39

/IR

RCCx39

/XR

Inner
diameter

mm

7.7

13.4

22.1

37.2

55.1

82.50

Wall thick-
ness

mm

0.89

1.24

1.65

2.6

2.9

3.2

pressure Equipment directive 97/23/EC

- Module : H; Fluid group : 1; Category : III

RCCx34-RCCx38

: Fluid group 2, SEP

RCCx39-RCCx39/XR

: Fluid group 2, Cat. I

For all process Connections: CRN 0F12074.5

Maximum tube pressure for SL/SH up to 27°C (RT=Room Temp.):
- RCCS30 / 31 / 32

: 285 bar

- RCCS33

: 185 bar

- RCCS34 / RCCT34

: 260 bar

- RCCS36 / RCCT36

: 210 bar

- RCCS38 / RCCT38 : 175 bar
- RCCS39 / RCCT39

: 135 bar

- RCCS39/IR / RCCT39/IR : 110 bar
- RCCS39/XR / RCCT39/XR: 95 bar
For higher medium temperatures maximum tube pressure
needs to be derated as follows :
up to 50 °C

:

4% derating

51 to 100 °C

:

11% derating

101 to 150 °C

:

20% derating

151 to 230 °C

:

30% derating

231 to 350 °C

:

38% derating

Higher pressure on request.
The maximum process pressure of a single instrument is
given by the lower value either of the process connections
(table 9) or tubes. The maximum temperature and process
pressure limits of an instrument are marked on the name-
plate as TS and PS.
The given temperature / pressure ranges are calculated and
approved without corrosion or erosion. The customer is fully
responsible to select proper materials to withstand his
corrosive or erosive conditions. In case of heavy corrosion
and/or erosion the instrument may not withstand the pres-
sure and an incident may happen with human and/or
environmental harm. Yokogawa will not take any liability
regarding damage caused by corrosion / erosion. If corrosion/
erosion may happen, the user has to check periodically if the
necessary wall thickness is still in place.

gas Content limits for liquid/gas Mixtures
Gas content limit is defined as the amount of gas in a liq-
uid/gas mixture which generates an error (frequency error) in
the converter. The gas content limit is dependent on viscosity,
surface tension and bubble size of the liquid/gas mixture.
Furthermore it is highly flow rate dependent (the higher the
flow rate, the lower the gas content limits). The stated values
are for a flow of 50% of Qnom and water/air without /HP:

- RCCS30 to 32

: no limitation

- RCCS33 non-Ex type

: no limitation

- RCCS33 Ex type

: approx. 35%

- RCCx34

: no limitation

- RCCx36

: approx. 50%

- RCCx38

: approx. 30%

- RCCx39

: approx. 7%

- RCCx39/IR

: approx. 3%

- RCCx39/XR

: approx. 2% (with option /HP)

With option /HP the gas content limits are improved.
With liquid/gas mixtures the specified mass flow accuracy will
not be achieved.
For short time aeration a function can be activated to keep
the current outputs constant during the aeration time.

Secondary Containment
The housings of the RCCS30-33 and the RCCx39/XR are
not rated for secondary containment. Rupture pressure for
RCCx34-38 is typical above 120bar, for RCCx39 above 80
bar, for RCCx39IR above 50bar.
However if the detector housing is exposed to this pressure
it will deform and measurement will be strongly influenced.
Therefore the pressure test of the housing (option /J1) can
only be done at the pressure where deformation does not
happen.