10 measurement system and applications, 10 measurement system and applications -8 – Yokogawa RotaMASS 3-Series User Manual
Page 26

3. product description
3-8
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
3.10 Measurement system and applications
ROTAMASS measures the mass flow of fluids directly.
The measurement system uses the Coriolis principle and is suitable for a wide range of continuous flow
measurement applications in all branches of process technology.
ROTAMASS has two components: the detector and the converter.
The detector measures the mass flow directly and converts it into electrical signals.
The converter evaluates the electrical signals and outputs the following values:
-
mass flow, independent from media properties, such as density, temperature, viscosity
-
fluid density
-
fluid temperature
The values are displayed and output as electrical values for use by other systems.
The converter is operated by three infra-red keys and a 4- line display and is standard equipped with HART-
communication protocol.
ROTAMASS is suitable for
-
measuring liquids, liquids with solid content, multi-phase mixtures
-
measuring gases (restricted by density and pressure loss)
-
simultaneous measurement of mass flow, density, temperature, volume flow and
cumulated mass and volume
-
measuring of concentrations of liquid mixtures, solutions and suspensions
-
connection to controllers and process control systems
ROTAMASS provides the following I/O-connections and can be configured for a wide
variety of different measurement applications (controlling, checking, monitoring, metering, mixing, filling).
-
2 active analog outputs
-
2 passive pulse outputs / status outputs
-
1 status input
optional /AP:
-
1 active pulse output / status output
optional /NM:
-
1 pulse output / status output according EN 50227 (NAMUR)
optional /KF2 :
-
1 passive analog output Ex ia
-
1 passive pulse output / status output Ex ia
These capabilities make ROTAMASS ideal for the increasing demand of requirements for automation and
the growing trend towards batch processes.
optional /FB see IM 01R04B05-00E-E :
-
1 Foundation Fieldbus communication line
optional /FB and /KF4 see IM 01R04B05-00E-E :
-
1 intrinsic safe Ex ia Foundation Fieldbus communication line