4. INSTAllATION

4-5

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

'BJ

1JQFGMBOHF

#PMU

/VU

(BTLFU

'MPXNFUFS'MBOHF

(3) Mounting Clamp
Clamp and gasket must be mounted to fit into the groove of the ferrules of the pipe, fitted to cover the taper
portion of the ferrule to the Coriolis flowmeter.
The ferrules, clamps, gaskets are not included and must be provided on site.

F0306.ai

Ferrule

Gasket

Clamp

gas flow measurement

A stable zero is mandatory for a good mass flow measurement. Mechanical installation stress and flow
noise influence zero stability. Action has to be taken to avoid any generation of sound.

Recommendations:
• Support the weight of the detector by soft coupling (silicon or other kind of rubber support.
• Do not bend or stress the detector via the adjacent pipe. This is achieved by supporting the pipe 10D or

more away from the detector.

• Pipe reduction or extension should be avoided directly before or after the meter.
• Avoid any control valves or orifices or any other sound generator near the detector.

CAUTION