2 setting temperature 20-120°c to analog output 2 – Yokogawa RotaMASS 3-Series User Manual

Page 58

Advertising
background image

5. BASIC OpErATINg prOCEdUrES

5-6

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

5.4.2 Setting Temperature 20-120°C to Analog Output 2

MFR

DEN

TMP

FTL

SET > 2s

Setting Enable?

Main Menu

Language

Process variables

Diag/Service

Main Menu

Basic Setup

Detailed Setup

Review device info

SET

SET

Basic Setup

Tag

Long Tag

Display config

Basic Setup

Temperature

Analog 1 select

Analog 2 select

5x

Temperature

Temperature unit

Temperature LRV

Temperature URV

SET

Temperature

Temperature LRV

Temperature URV

Temperature damping

SET

Temperature LRV

000.00 degC

000.00 degC

Edit number with SHIFT and

Temperature LRV

000.00 degC

020.00 degC

2x SET

Temperature

Temperature URV

Temperature damping

Temperature

Temperature LRV

Temperature URV

Temperature damping

F56.EPS

3x

123.450 kg/h

0.99700 kg/l

12.3400 degC

12.3450 kg

Temperature unit