7. pArAMETEr dESCrIpTION
7-18
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
parameter
data
form
data range
dec.
point
Unit
r : read
W : write
Exe :
execute
reset
Master
reset
power on
default
Concentration meas format Select
xxxxxxx
-
-
R/W
X
xxxx.XX
xxxxx.X
X
xxxx.XX
X
xxx.XXX
X
xx.XXXX
X
x.XXXXX
X
Concentration meas LRV
Decimal
0 to 110
0 to 5
Concentration
meas unit
R/W
X
0
Concentration meas URV
Decimal
0 to 110
0 to 5
Concentration
meas unit
R/W
X
100
Concentration meas Damp
Decimal
0 to 200
1
Sec
R/W
X
10
Concentration meas lowcut Decimal
0 to 10%
0 to 5
Concentration
meas unit
R/W
0
Conc meas alm 1 sel
Select
No function
-
-
R/W
X
No function
More than
Less than
Conc meas alm 1 crit
Decimal
0 to 110% of URV
0 to 5
Concentration
meas unit
R/W
X
LRV
Conc meas alm 2 sel
Select
No function
-
-
R/W
X
No function
More than
Less than
Conc meas alm 2 crit
Decimal
0 to 110% of URV
0 to 5
Concentration
meas unit
R/W
X
URV
Net flow select
Select
Mass
-
-
R/W
X
Mass
Volume