5 output trim, 6 detector cleaning, 5 output trim -4 – Yokogawa RotaMASS 3-Series User Manual

Page 180: 6 detector cleaning -4, 5 output trim 8.6 detector cleaning

Advertising
background image

8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-4

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.5 Output trim

8.6 detector cleaning

CAUTION

For integral type : Ambient temperature may not exceed 55°C !

The equipment can be CIP-cleaned. Steam below 230°C is not a problem. However, you should take care
that the cleaning operation does not last so long that the converter temperature exceed 55°C.
Switch power off during cleaning.

Output trim

With this function the 4 mA and the 20 mA can be adjusted if
required

AO 1 trim 4mA / AO 1 trim 20mA

The analog output 1 is set to the 4 mA or 20 mA. Connect a load
resistor between 20 and 600 W and measure the output current.
This current value must be entered on request on display.
The deviation to 4 mA is corrected by the converter itself.

AO 2 trim 4mA / AO 2 trim 20mA

The analog output 2 is set to the internal 4 mA or 20 mA value.
Connect a load resistor between 20 and 600 W and measure the
output current.
This current value must be entered on request on display.

Advertising