Installation, 1 installation conditions, Danger caution – Yokogawa IR400 User Manual

Page 14

Advertising
background image

3 - 1

IM 11G02N01-01E

This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent

explosion, fire or other serious accidents.

• Entrust the installation, movement or re-installation to a specialist or the supplier. A poor

installation may cause accidental tipover, shock hazard, fire, injury, etc.

• The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tip

over or drop, for example, causing accident or injury.

• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an

injury.

• This unit should be installed in a place which conforms to the conditions noted in the

instruction manual. Otherwise, it may cause electric shocks, fire or malfunction of the unit.

• During installation work, care should be taken to keep the unit free from entry of cable chips

or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.

DANGER

CAUTION

3. INSTALLATION

3.1 Installation conditions

To install the analyzer for optimum performance, select a location that meets the following
conditions;
(1) This instrument is system built in type. This instrument should be used while embedded
in a panel, locker, or enclosure of steel sheet.
(2) Use this instrument indoors.
(3) A vibration-free place
(4) A place which is clean around the analyzer.
(5) Power supply
Rated voltage

: 100 V to 240 V AC

Operating voltage

: 85 V to 264 V AC

Rated frequency

: 50/60 Hz

Power consumption

: 250 VA max.

Inlet

: Comformity to EN60320 class I type 3-pin inlet

(6) Operation conditions
Ambient temperature

: -5˚ to 45˚

C

Ambient humidity

: 90% RH or less, no condensation